Sarnatskiy, A (2016) Calculation of stopping powers for protons using single point calibration method for dual energy CT. Bachelor's Thesis, Physics.
|
Text
A_Sarnatskiy_E_Biegun_E_van_de_1.pdf - Published Version Download (853kB) | Preview |
|
Text
Toestemming.pdf - Other Restricted to Backend only Download (497kB) |
Abstract
The single point calibration method introduced by Hünemohr in 2014 is analysed in this study for DECT energies 90/150Sn kV. The composition of 38 calibration materials have been used to investigate, which predicts the relative electron number density and effective atomic number accurately. Also an optimal constant α has been determined to be 3.3. The best calibration materials have been used in calculating the relative stopping power for 81 human tissues. The calibration materials AlMgSi, Al2O3, Carbon, SB3, Methanol show less than 1 percent relative difference compared to the relative stopping power calculated from composition; KCl even showed less than 0.1 percent accuracy. This was a theoretical study, so an experimental one is needed to confirm these results.
Item Type: | Thesis (Bachelor's Thesis) |
---|---|
Degree programme: | Physics |
Thesis type: | Bachelor's Thesis |
Language: | English |
Date Deposited: | 15 Feb 2018 08:24 |
Last Modified: | 15 Feb 2018 08:24 |
URI: | https://fse.studenttheses.ub.rug.nl/id/eprint/14495 |