Javascript must be enabled for the correct page display

The quality of δ13CO2 measurements with the NDIR: ABB EL3020

Visser, Wietse (2021) The quality of δ13CO2 measurements with the NDIR: ABB EL3020. Bachelor's Thesis, Physics.

[img]
Preview
Text
bachelorthesis Wietse Visser the_Realiability_of_C_13_measurements_with_an_ABB_EL3020.pdf

Download (1MB) | Preview
[img] Text
toestemming.pdf
Restricted to Registered users only

Download (97kB)

Abstract

In this article the quality of the NDIR: ABB EL3020 is tested. Between 20 October and 1 November 2020 the CO2 and 13CO2 levels were measured in Groningen, The Netherlands. When measured accurately δ 13CO2 can be useful for source attribution of CO2 emissions. After calibration, the CO2, 13CO2 and, δ 13CO2 levels were calculated. The reliability and accuracy of the CO2 and 13CO2 levels were reasonable. The accuracy of the δ 13CO2 was also reasonable, with a standard deviation of ±0.26h, but the δ 13CO2 level was not reliable. The spread of the data was ±2h, while the differences between δ 13CO2 values should typically be within one per mille and should be able to follow with a precision of less than 0.1h on an hourly basis. Even when the data was averaged out the spread stayed too high to be useful

Item Type: Thesis (Bachelor's Thesis)
Supervisor name: Meijer, H.A.J.
Degree programme: Physics
Thesis type: Bachelor's Thesis
Language: English
Date Deposited: 18 Jan 2021 10:29
Last Modified: 18 Jan 2021 10:31
URI: https://fse.studenttheses.ub.rug.nl/id/eprint/23813

Actions (login required)

View Item View Item